Optiflow™ Junior 2 Introduction and Fitting Guide | F&P Healthcare
FULL TRANSCRIPT
[Music]
Fisher and Paykel Healthcare opt afloat
Junior 2 is a nasal canula specifically
designed for the delicate anatomy and
flow requirements of your smallest
patients on nasal high flow therapy it
uses new waveflex technology for
enhanced prong retention and offer sizes
to fit a wide range of patients while
retaining the original benefits of opti
flow junior it's the next generation of
care up to flow junior to has five sizes
in its range our newest extra small
cannula released for the smallest of
patients the purpose of the in-service
video is to explain how to correctly fit
and use the up to flow junior to nasal
cannula this in-service video has been
split into three parts fitting cannula
removing cannula replacing cannula
important before application of the
cannula follow your hospital protocol
for skin preparation and gas source
setup
fitting the up to Flo Junior to cannula
select appropriate cannula size
recommended narrow occlusion is
approximately 50% patient weight should
only be used as a guide the infant snare
should be the primary consideration
choose appropriate cannula size in this
case the patient is a medium before
application of the cannula follow your
hospital protocol for skin preparation
the medium sized cannula selected can
now be removed from the packaging
connect to the fisher & paykel
healthcare circuit check the connection
by feeling the air flow through the
prongs the circuit can be secured to the
bedding using the tubing clip remove the
first backing tabs from the wiggle pads
to and avoid touching the adhesive keep
the second backing tabs in place insert
the cannula into the nares ensure the
cannula bridge rest close to the nose
without touching the septum do not
stretch the cannula during application
stick the wiggle pads to to the
patient's cheeks remove the second
backing tabs and stick the wiggle pads
to onto the cheeks one at a time
opted to flow jr. to nasal canula checks
ensure the recommend and narrow clusion
is approximately 50% ensure the cannula
bridge rest close to the nose without
touching the septum ensure the patient
does not lie on the tubing as this may
apply pressure to the patient's ears or
face check summary check one size and
space check to sensitive septum check
three tidy tubing readjust if required
removing the up to flow jr. to canula
place fingertip on the outside edge of
the wiggle pads to and gently peel the
cannula away from the wiggle pads to
starting from the outside peel towards
the nose replacing the wiggle pads to
remove the up to flow jr. to cannula use
a damp cloth to wipe the patients skin
and the underside of the wiggle pads too
while gently peeling away from the
patient's face adhere the replacement
wiggle pads to to the cannula remove the
first backing tabs and stick the wiggle
pads to onto the patient's cheeks
remove the second backing tabs and stick
onto the patient's cheeks
[Music]
warning appropriate patient monitoring
eg oxygen saturation must be used at all
times failure to monitor the patient eg
in the event of an interruption to gas
flow may result in serious harm or death
we hope you have enjoyed this in-service
video on the fisher & paykel healthcare
up to flow junior to nasal cannula
please contact your local representative
for further information about these or
any other fisher & paykel health care
products or therapies or visit WWF ph
care comm
you
[Music]
UNLOCK MORE
Sign up free to access premium features
INTERACTIVE VIEWER
Watch the video with synced subtitles, adjustable overlay, and full playback control.
AI SUMMARY
Get an instant AI-generated summary of the video content, key points, and takeaways.
TRANSLATE
Translate the transcript to 100+ languages with one click. Download in any format.
MIND MAP
Visualize the transcript as an interactive mind map. Understand structure at a glance.
CHAT WITH TRANSCRIPT
Ask questions about the video content. Get answers powered by AI directly from the transcript.
GET MORE FROM YOUR TRANSCRIPTS
Sign up for free and unlock interactive viewer, AI summaries, translations, mind maps, and more. No credit card required.